Troubleshooting
This section presents tables of technical alarm and information messages, as well as
problem descriptions that do not generate messages, to help you troubleshoot issues on
the monitor.
Note
Problem descriptions without messages appear at the end of this section.
When the monitor detects certain events, a message appears in the Device Status area
at the top of the screen. Message types include the following:
•
Information messages, which appear on a blue background.
•
Very low-priority alarms, which appear on a cyan background.
•
Low- and medium-priority alarms, which appear on an amber background.
•
High-priority alarms, which appear on a red background.
Technical alarm messages are low or very low priority unless noted in the Message
column.
You can dismiss a message by touching the message on the screen, or, for some
messages, you can wait for the message to time out.
To use these tables, locate the message that displays on the monitor in the left column
of the table. The remainder of the row explains possible causes and suggests actions
that can resolve the issue.
Note
Instructions to "Call for service" in the following tables mean that you
should contact qualified service personnel in your facility to investigate the
issue.
Patient movement messages
Message
Possible cause
Suggested action
Replace the bed sensor.
The sensor is faulty or expired
Replace the bed sensor.
The cable is faulty or expired
Replace the cable.
The bed sensor is disconnected.
The bed sensor is disconnected from the
monitor
Check the bed sensor cable to
ensure that connections to the
monitor and the extension cable
are tight.
The bed sensor is disconnected from the
extension cable
207
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......