Attach the temperature probe
CAUTION
The temperature module only operates with the probe well
properly in place.
1. Remove the cover of the temperature module by pressing the tab on the bottom and
sliding the cover to the right. The cover is located at the bottom-right of the monitor,
below the probe well.
2. Hold the temperature probe cable connector with the spring tab on the right and
insert it into the probe port of the temperature module.
3. Push it into place until it clicks.
4. Reattach the cover. Use the alignment tab and slide the cover to the left to click it
back into place.
Remove the temperature probe and well
Follow these steps to disconnect the probe cable and remove the probe well.
1. Remove the cover of the temperature module by pressing the tab on the bottom and
sliding the cover to the right. The cover is located at the bottom-right of the monitor,
below the probe well.
34 Setup
Welch Allyn Connex
®
Devices
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......