Message
Possible cause
Suggested action
RRa background interference
detected.
Excessive ambient or environmental noise
Reduce the level of noise in the
room.
NIBP messages
Message
Possible cause
Suggested action
NIBP air leak; check cuff and tubing
connections.
The NIBP module has an air leak
Check the cuff, tubing, and
connections for leaks. If no leaks
are found, clear the alarm and
retry NIBP.
If the message reappears, call for
service to replace the NIBP
module.
NIBP not functional. Call for
service.
A module error occurred
Call for service.
The ambient temperature is out of range
Use the monitor in the specified
temperature range.
Unable to determine NIBP; check
connections; limit patient
movement.
The NIBP module detected a motion artifact
Check connections; limit patient
movement.
Clear the alarm and retry NIBP.
Unable to determine NIBP; check
connections and tubing for kinks.
The NIBP tubing on the outside of the device has
a kink
Check the connections and tubing
for kinks.
Clear the alarm and retry NIBP.
The NIBP module needs to be calibrated
Call for service to calibrate the
NIBP module.
Tubing inside the NIBP module is kinked
Call for service to replace the
NIBP module.
Incorrect NIBP cuff size; check
patient type.
The cuff size is not correct for the selected
patient type
Check the patient type and cuff
size.
Clear the alarm and retry NIBP.
Inflation too quick; check NIBP cuff
and tubing connections.
NIBP inflation was too quick
Check the connections and tubing
for kinks.
Clear the alarm and retry NIBP.
Unable to determine NIBP; check
inflation settings.
Target pressure is too low
Check inflation settings and
change as necessary.
Clear the alarm and retry NIBP.
Change the cuff inflation target
(CIT).
Excessive patient movement.
The NIBP module detected a motion artifact
Touch
OK
to dismiss.
Limit patient movement and retry
NIBP.
Directions for use
Troubleshooting 211
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......