Note
Event data (e.g., physiological alarms, patient movement,
manual NIBP or temperature measurements, Manual
parameter entries) is displayed in all trend intervals. By
modifying the trend interval to one of the longer time spans,
you can filter out non-alarming, continuous measurements
between the currently selected time intervals to focus more
attention on event data. Alternatively, you can modify the
trend interval by selecting a shorter time span to display a
more comprehensive listing of continuous measurements.
3. Use the scroll bar on the right to view patient measurements for any parameters that
do not fit in the display area.
4. Touch the page forward
and page backward
controls in the table header
to view additional measurements for this patient. The most recent measurements
appear on the right of the table, with older measurements to the left.
5. Touch the graphical trends button
to view a graphical representation of
patient measurements recorded in the tabular trends table.
6. Touch the tabular trends button
to return to the tabular view.
7. Touch
to print the patient's record.
A Print options dialog appears.
8. Select the desired Timespan and touch
.
Note
If no patient is selected, the Continuous Monitoring profile is not active,
and no measurements have been taken, all table cells in the Review tab
remain blank.
Note
In the Continuous Monitoring profile, you cannot delete saved patient
measurements. Patient measurements older than 24 hours are
automatically deleted from the Review tab.
Note
The date and time stamps on saved patient measurements will adjust in
response to new date and time settings.
96 Patient data management
Welch Allyn Connex
®
Devices
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......