parameter. See "Configure movement alarms" to adjust alarm settings. Follow these
steps to set up movement modifiers, if applicable.
1. Touch the
Patients
tab.
2. Touch the
Manual
tab.
3. Scroll through the list to find the Movement section, then enter or select the
Movement modifier(s) as desired.
Note
If no modifiers are associated with this parameter, no
Movement section will appear.
4. Touch
OK
.
The Home tab appears. In the Continuous Monitoring profile, the modifiers are saved
with the next set of measurements sent to the network.
Configure movement alarms and patient turn reminder
Follow these steps to set alarm limits for patient movement.
1. Verify that you are using the Continuous Monitoring profile.
Note
If “Allow profile change” is enabled in Advanced settings,
the device attempts to shift automatically to the Continuous
Monitoring profile when you connect the sensor cable.
2. Touch the
Alarms
tab.
3. Touch the
Movement
vertical tab.
4. Ensure that the Exit alarm and Motion alarm controls are set to ON.
Note
If any parameter's alarm limit control is set to OFF, you
cannot adjust alarm limits on the Alarm tab, and no visual or
audio signals occur for that specific parameter.
5. Adjust the following settings as desired:
•
Exit sensitivity
. Use the up/down arrow keys or the keypad to choose a
sensitivity level ranging from 1 to 6. Setting 1 is the least sensitive, and setting 6
is the most sensitive.
Note
Levels 5 and 6 are highly sensitive and might trigger an
alarm as a result of minimal movement while the patient
is in bed. Correct patient assessment is the key to
minimize false alerts.
•
Turn reminder
. Select the desired turn schedule from the list box.
6. Touch the
Home
tab.
The new alarm settings take effect immediately.
Monitor patient movement
WARNING
Patient safety risk. Do not use heating blankets. Heating
blankets could reduce the safety and efficacy of the system and void the
warranty.
118 Patient monitoring
Welch Allyn Connex
®
Devices
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......