Configure SpHb alarms
Follow these steps to set alarm limits for SpHb measurements.
1. Verify that you are using the Intervals Monitoring or Continuous Monitoring profile.
2. Touch the
Alarms
tab.
3. Touch the
SpHb
vertical tab.
4. Ensure that the SpHb alarm limit control is set to ON.
Note
If any parameter's alarm limit control is set to OFF, you
cannot adjust alarm limits on the Alarm tab, and no visual or
audio signals will occur for that specific parameter.
5. Enter the desired upper and lower alarm limits for SpHb using the up/down arrow
keys or the keypad.
6. Touch the
Home
tab.
The new alarm settings take effect immediately.
Set SpHb averaging mode
Touch
in the SpHb frame.
The SpHb frame displays the current mode.
Measure SpHb
WARNING
Inaccurate measurement risk. Use only Masimo rainbow SET
sensors and accessories on Masimo-equipped monitors.
WARNING
High-intensity lights (including pulsating strobe lights) directed
on the pulse co-oximeter can prevent it from obtaining readings.
WARNING
The pulsations from intra-aortic balloon support can increase
the pulse rate displayed on the monitor. Verify the patient's pulse rate
against the ECG heart rate.
WARNING
Inaccurate measurement risk. Venous congestion might cause
a low reading of arterial oxygen saturation. Place the sensor on a hand at
heart level to ensure proper venous outflow from a monitored site.
WARNING
Inaccurate measurement risk. The pulse co-oximeter can be
used during defibrillation, but the reading may be inaccurate for up to 20
seconds.
WARNING
Patient injury risk. Do not attempt to reprocess, recondition, or
modify any sensors or patient cables. Doing so might damage electrical
components.
WARNING
Pulse rate measurement might not detect certain arrhythmias
because it is based on the optical detection of a peripheral flow pulse. Do
not use the pulse oximeter as a replacement or substitute for ECG-based
arrhythmia analysis.
Directions for use
Patient monitoring 171
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......