Icons on the Home tab
Icons in parameter frames
The icons in the parameter frames indicate alarm notification settings. When alarm limits
are on, the icons are black and white until an alarm occurs. Then, the icons change color
to indicate the priority of the alarm. Red icons represent high-priority alarms, and amber
icons represent medium- or low-priority alarms.
Icons in parameter frames
Icon
Name and status
Alarm off.
No visual or audio alarms occur for this parameter.
You are fully responsible for maintaining awareness of the patient’s condition.
Alarm on.
Audio and visual notifications are enabled.
Alarm audio off.
Only visual notifications occur.
If the patient is being monitored continuously, maintain a clear line of sight with the monitor while
audio alarms are disabled.
Alarm audio paused.
The audio tone is paused. The icon remains until the paused time counts down to 0.
Icons in the Device Status area
The icons in the Device Status area are black and white, but the background area
changes colors to indicate the alarm priority. Messages accompany these icons. These
icons can be controls or status indicators.
Icons in the Device Status area
Icon
Name and status
Alarm active.
One or more alarms are active. Touch this icon to reset (pause or turn off) the alarm tone.
WARNING
Do not pause or turn off an audible alarm if patient safety might be
compromised.
Alarm audio off.
Audio signals are disabled, but alarm limits and visual alarm signals remain active.
If the patient is being monitored continuously, maintain a clear line of sight with the monitor while
audio alarms are disabled.
104 Alarms
Welch Allyn Connex
®
Devices
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......