The frame provides a numeric view and a graphical trend view of IPI data. You can toggle
between views by touching the left side of the frame.
IPI numeric view
This view displays the numeric value for the integrated parameters.
The numeric view also includes a graphic indicator of the patient's current IPI value. As
shown in the examples below, when the IPI value is 8, 9, or 10, the outline of the vertical
bar is gray. When the IPI value is 4, 5, 6, or 7, the outline of the bar is yellow, and the
middle section of the bar is shaded yellow. When the IPI value is 1, 2, or 3, the outline of
the bar is red, and the lowest section of the bar is shaded red.
IPI trend graph view
This view presents a graph of the IPI numeric values over a user-selected period and
thus can alert you to changes in the patient's respiratory status. In Advanced settings,
you can select the period displayed.
The size of the IPI frame and displayed trend data varies based on your configuration.
The graph shows the IPI value on the y-axis and time on the x-axis (oldest calculations on
the left and newest calculations on the right). IPI values update once every second.
Set up IPI
Follow these steps to configure IPI parameter options.
1. Touch the
Settings
tab.
2. Touch the
Setup
tab.
3. Touch the
IPI
vertical tab.
4. Adjust the Trend period as desired. You can set the default in Advanced settings.
5. Touch the
Home
tab.
The new settings take effect immediately.
6. Touch the patient type control on the right side of the Patient frame.
The Patient Summary tab appears.
Directions for use
Patient monitoring 129
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......