7. Connect the sampling line to the patient as described in the directions for use
supplied with the sampling line.
8. Perform the following checks:
a. Check that CO2 values appear on the monitor.
b. Check the patient type. Make sure it is correct to ensure accurate sampling.
c. Touch the alarm control to ensure the alarms are set appropriately.
d. Toggle to the waveform view to ensure that a complete waveform appears.
Note
When the cannula is not on the patient, the etCO2 value is blank.
Note
After you connect the sampling line, the monitor immediately searches for
breaths, but it does not indicate a No Breath condition before any valid
breaths have been detected.
Note
To avoid moisture buildup and sampling line occlusion during suction or
nebulization of intubated patients, stop the CO2 pump and remove the
sampling line luer connector from the monitor.
Note
Replace the sampling line according to hospital protocol or when a
blockage is indicated by the device. Excessive patient secretions or a
buildup of liquids in the airway tubing might occlude the sampling line,
requiring more frequent replacement.
Note
When the message "Check for occlusion in gas line" appears on the screen,
indicating that the sampling line attached to the monitor is blocked, the
monitor’s CO2 pump will stop. Follow the instructions in the
Troubleshooting section of this manual to correct the problem.
Stop and restart the CO2 pump
Stopping the CO2 pump prevents the ingress of fluids into the CO2 module during
certain patient procedures (such as suction or lavage). In addition, this pump-stop
condition pauses the monitoring of all parameters obtained from the CO2 module for a
configured period of time or until you restart the CO2 pump.
1. To temporarily pause or stop CO2 monitoring, touch the pump-stop button
in the etCO2 frame.
Note
This button is a toggle. When you stop the CO2 pump, the
Start button with a countdown timer appears.
2. To resume CO2 monitoring, touch the pump-start button
.
The CO2 pump restarts, the current timer clears, and the Stop button appears in the
frame.
Respiration rate
RR frame
The RR frame displays the following:
126 Patient monitoring
Welch Allyn Connex
®
Devices
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......