4. Specify settings.
Setting
Action/Description
Automatically send on manual save
Select this option to specify that measurements are sent
to the network when a clinician saves measurements on
the Home tab.
Note
When the device is in the Continuous Monitoring
profile, this setting is grayed out.
Note
When the monitor is not connected to the network,
measurements saved on the monitor are sent to the
network with the first successful send after reconnecting
to the network.
Delete readings after successful send
Select this option to specify that measurements are
deleted from the monitor after they are successfully sent
to the network. Sent measurements do not appear in the
Review tab.
Emulate Spot Vital Signs LXi
Select this option to specify that clinical data sent to the
network appears as Spot Vital Signs LXi data at the
network.
Connect to CS
Select to enable connection to the Connex central
station.
Enable clear button on Home tab
Select to enable the Clear button on the Home tab in
Intervals Monitoring, Office, and Spot Check profiles.
When not selected (disabled), the Clear button does not
appear on the Home tab.
5. Do one of the following:
•
To continue in Advanced Settings, touch another tab.
•
To exit the Advanced Settings and return to the Home tab, touch
Exit
.
Network
View advanced monitor information
The Status tab shows the monitor's software version, MAC and IP addresses, network,
server and access point information, session information, and more.
1. Access the Advanced Settings.
a. Touch the
Settings
tab.
b. Touch the
Advanced
tab.
c. Enter the
Advanced settings code
.
d. Touch
OK
.
The General tab appears.
2. Touch the
Network
tab.
3. Touch the
Status
tab.
4. View the information.
5. Do one of the following:
202 Advanced settings
Welch Allyn Connex
®
Devices
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......