Alarm limits
Upper limit range of entry
Lower limit range of entry
Pediatric: 17 to 140 mmHg (2.3 to 18.7 kPa).
Factory default: 110 mmHg (14.7 kPa).
Neonate: 12 to 110 mmHg (1.6 to 14.7 kPa).
Factory default: 80 mmHg (10.7 kPa).
Pediatric: 15 to 138 mmHg (2.0 to 18.4 kPa).
Factory default: 50 mmHg (6.7 kPa).
Neonate: 10 to 108 mmHg (1.3 to 14.4 kPa).
Factory default: 35 mmHg (4.7 kPa).
SpO2
Adult, Pediatric, Neonate: 52% to 100%. Factory
default: 100%.
Adult, Pediatric, Neonate: 50% to 98%. Factory
default: 90%.
SpHb
Adult, Pediatric, Neonate: 1.5 to 24.5 g/dL (1.5
to 15.0 mmol/L). Factory default: 17.0 g/dL (11.0
mmol/L).
Adult, Pediatric, Neonate: 1.0 to 24.0 g/dL (1.0
to 14.5 mmol/L). Factory default: 7.0 g/dL (4.0
mmol/L).
Pulse rate (NIBP,
SpO2)
Adult: 27 to 300 beats per minute. Factory
default: 120 beats per minute.
Pediatric: 27 to 300 beats per minute. Factory
default: 150 beats per minute.
Neonate: 27 to 300 beats per minute. Factory
default: 200 beats per minute.
Adult: 25 to 298 beats per minute. Factory
default: 50 beats per minute.
Pediatric: 25 to 298 beats per minute. Factory
default: 50 beats per minute.
Neonate: 25 to 298 beats per minute. Factory
default: 100 beats per minute.
Pulse rate
(EarlySense)
Adult, Pediatric: 37 to 150 beats per minute.
Factory default: 130 beats per minute.
Adult: 35 to 148 beats per minute. Factory
default: 40 beats per minute.
Temperature
Adult, Pediatric, Neonate: 87.1 to 110.0°F (30.6
to 43.3°C). Factory default: 101.0°F (38.3°C).
Adult, Pediatric, Neonate: 85.1 to 108.0°F (29.5
to 42.2°C). Factory default: 94.0°F (34.4°C).
etCO2
Adult: 2 to 150 mmHg (0.3 to 20.0 kPa). Factory
default: 60 mmHg (8.0 kPa).
Pediatric: 2 to 150 mmHg (0.3 to 20.0 kPa).
Factory default: 60 mmHg (8.0 kPa).
Neonate: 2 to 150 mmHg (0.3 to 20.0 kPa).
Factory default: 50 mmHg (6.7 kPa).
Adult: 0 to 148 mmHg (0.0 to 19.7 kPa). Factory
default: 15 mmHg (2.0 kPa).
Pediatric: 0 to 148 mmHg (0.0 to 19.7 kPa).
Factory default: 15 mmHg (2.0 kPa).
Neonate: 0 to 148 mmHg (0.0 to 19.7 kPa).
Factory default: 20 mmHg (2.7 kPa).
FiCO2
Adult: 0 to 150 mmHg (0.0 to 20.0 kPa). Factory
default: 8 mmHg (1.1 kPa).
Pediatric: 0 to 150 mmHg (0.0 to 20.0 kPa).
Factory default: 8 mmHg (1.1 kPa).
Neonate: 0 to 150 mmHg (0.0 to 20.0 kPa).
Factory default: 5 mmHg (0.7 kPa).
None
IPI
None
Adult, Pediatric: 1 to 10. Factory default: 4.
RR (CO2)
Adult: 5 to 150 breaths per minute. Factory
default: 50 breaths per minute.
Pediatric: 5 to 150 breaths per minute. Factory
default: 50 breaths per minute.
Neonate: 5 to 150 breaths per minute. Factory
default: 80 breaths per minute.
Adult: 0 to 145 breaths per minute. Factory
default: 3 breaths per minute.
Pediatric: 0 to 145 breaths per minute. Factory
default: 3 breaths per minute.
Neonate: 0 to 145 breaths per minute. Factory
default: 12 breaths per minute.
RR (EarlySense)
Adult, Pediatric: 9 to 44 breaths per minute.
Factory default: 32 breaths per minute.
Adult, Pediatric: 8 to 43 breaths per minute.
Factory default: 8 breaths per minute.
234 Specifications
Welch Allyn Connex
®
Devices
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......