Alarm limits
Upper limit range of entry
Lower limit range of entry
RRa
Adult: 5 to 150 breaths per minute. Factory
default: 50 breaths per minute.
Adult: 0 to 145 breaths per minute. Factory
default: 3 breaths per minute.
Patient movement
Exit sensitivity: 1 to 6. Factory default: 3
EH > 1 minute
Off
Alarm delays
Physiological parameter
Default delay
(seconds)
Configurable
range
(seconds)
1
Sum of default
delays
(seconds)
2
Sum of
configurable
delays
(seconds)
Pulse rate
SpO2
3
3–120
4
4–121
Motion (EarlySense)
3
3–120
4
4–121
SpO2
10
0–30
11
1–31
SpHb
10
0–10
11
1–11
etCO2
15
0–30
16
1–31
etCO2 no breath detected
Adult: 30
10–60
31
11–61
Pediatric: 20
10–60
21
11–61
Neonate: 15
10–60
16
11–61
Motion low confidence RR/PR
alarm delay
180
180–1800
181
181–1801
Respiration rate
RRa (Masimo)
10
0–60
11
1–61
CO2
10
0–60
11
1–61
Motion (EarlySense)
180
10–180
181
11–181
1
These delays are configurable in Advanced settings.
2
When the device determines the alarm condition, it generates alarm signals (visual and auditory) and notifies any
system to which it is connected within one second.
Note
The maximum alarm signal generation delay of a technical alarm when connected to a central station
is 4 seconds.
Note
There are no inherent alarm condition delays in the system for temperature or blood pressure alarms.
Consider ALL of these alarm delays when setting up your monitor and remote alarm system.
Directions for use
Specifications 235
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......