Nurse Call connection specifications
Nurse Call
50V DC maximum at 500mA (DC or ACRMS)
Handle specifications
Handle output
3.00 - 3.90 V, .700 - 1.5 A
Leakage current is less than 10 microamps from any exposed metal part.
NIBP specifications
Units of measure
Systolic, diastolic, MAP: mmHg, kPa; user-selectable
Pulse rate: beats per minute
Cuff pressure range
Meets or exceeds ANSI/AAMI SP10:2002 standards for cuff pressure range
Systolic range
Adult: 30 to 260 mmHg (4.0 to 34.7 kPa) (StepBP, SureBP)
Pediatric: 30 to 260 mmHg (4.0 to 34.7 kPa) (StepBP, SureBP)
Neonate: 20 to 120 mmHg (2.7 to 16.0 kPa) (StepBP)
Diastolic range
Adult: 20 to 220 mmHg (2.7 to 29.3 kPa) (StepBP, SureBP)
Pediatric: 20 to 220 mmHg (2.7 to 29.3 kPa) (StepBP, SureBP)
Neonate: 10 to 110 mmHg (1.3 to 14.7 kPa) (StepBP)
Cuff Inflation Target
Adult: 160 mmHg (21.3 kPa) (StepBP)
Pediatric: 120 mmHg (16.0 kPa) (StepBP)
Neonate: 90 mmHg (12.0 kPa) (StepBP)
Maximum Target Pressure
Adult: 280 mmHg (37.3 kPa) (StepBP, SureBP)
Pediatric: 280 mmHg (37.3 kPa) (StepBP, SureBP)
Neonate: 130 mmHg (17.3 kPa) (StepBP)
Blood pressure determination time
Typical: 15 seconds
Maximum: 150 seconds
Blood pressure accuracy
Meets or exceeds ANSI/AAMI SP10:2002 standards for noninvasive blood
pressure accuracy, ±5 mmHg (0.7 kPa) mean error, 8 mmHg (1.1 kPa)
standard deviation
Mean Arterial Pressure (MAP) range
The formula used to calculate MAP yields
an approximate value.
Adult: 23 to 230 mmHg (3.1 to 30.7 kPa) (StepBP, SureBP)
Pediatric: 23 to 230 mmHg (3.1 to 30.7 kPa) (StepBP, SureBP)
Neonate: 13 to 110 mmHg (1.7 to 14.7 kPa) (StepBP)
Pulse rate range (using blood pressure
determination)
Adult: 30 to 200 beats per minute (StepBP, SureBP)
Pediatric: 30 to 200 beats per minute (StepBP, SureBP)
Neonate: 35 to 220 beats per minute (StepBP)
224 Specifications
Welch Allyn Connex
®
Devices
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......