background image

Questions? Comments?

1-800-363-2381 • www.amgmedical.com

A.M.G. Medical Inc.

Montréal, QC H4T 1V5 • 1-800-363-2381

West Chazy, NY 12992 • 1-888-412-4992

R

e

v5 

06

1

8  

P/N 

86

1-7

1

5430

Physiotherapy • Physiothérapie

Instructions

Please read this manual thoroughly
before using this device for the first time.

Generated by : 

Jean-François Perron

Flat Size : 

24,811 cm x 18,81 cm

Assembled Size : 

folded: 10,5 x 15 cm 

                                open: 21 x 15 cm

Projet : 

715-430 TENS instructions EN rev5 0618

Item No. : 

715-430

    

Part No. : 

861-715430

Colours :     B/W             CMYK                  Spots:

Summary of Contents for 715-430

Page 1: ... N 861 715430 Physiotherapy Physiothérapie Instructions Please read this manual thoroughly before using this device for the first time Generated by Jean François Perron Flat Size 24 811 cm x 18 81 cm Assembled Size folded 10 5 x 15 cm open 21 x 15 cm Projet 715 430 TENS instructions EN rev5 0618 Item No 715 430 Part No 861 715430 Colours B W CMYK Spots ...

Page 2: ...EW 6 7 Components and parts Display symbols Function keys INSTRUCTIONS FOR USE 9 Complete step by step TO CREATE A CUSTOMIZED PROGRAM 12 PRE SET PROGRAMS 13 TENS programs EMS programs MASSAGE programs CLEANING AND MAINTENANCE 18 Cleaning the device and accessories Maintenance TROUBLESHOOTING 20 STORAGE 21 DISPOSAL 21 WARRANTY 22 TECHNICAL INFORMATION 22 ELECTRODE PLACEMENT GUIDE 23 BATTERY REPLACE...

Page 3: ...actions This can help improve and maintain muscle tone without actual physical activity and therefore prevent atrophy SAFETY INFORMATION 1 This device is recommended for Symptomatic relief of chronic intractable pain acute post traumatic pain or acute post surgical pain Improving local blood circulation Relief from muscle spasms Immediate post surgical stimulation of muscles to prevent venous thro...

Page 4: ...e skin irritation Use only as prescribed by a physician Never use in rooms where aerosols sprays are used or pure oxygen is being administered Do not use it near highly flammable substances gases or explosives Apply the electrodes to clean dry and unbroken skin only Keep electrodes separate during treatment Electrodes coming in contact with each other could result in improper stimulation or skin b...

Page 5: ...pply electrodes to the exact site indicated by your physician The electrodes should be placed firmly and evenly on the skin to ensure good contact See p 23 for our electrode placement guide which also includes recommendations for stimulation positions 9 8 Increases decreases the output intensity of channel 1 In setting mode press Ch1 to select the next parameter Turns the device on or off In treat...

Page 6: ...in treatment Use CH1 and CH2 to decrease the intensity The LCD will display the current intensity see figure III and remaining treatment time IMPORTANT The intensity of stimulation may be adjusted as required to a maximum of 90 1 level step Always turn the unit off before removing the electrodes The safety lock feature automatically activates after 20 seconds of button inactivity You cannot increa...

Page 7: ...en press CH2 and CH2 to adjust Press to confirm and go into standby then use CH1 and CH2 to adjust the amplitude of each channel NOTE to change the parameters press and hold P again PRESET PROGRAMS IN TENS MODE Note Treatment length for all options below is 30 minutes Synchronous 100 200 180 180 180 330 200 100 200 70 180 200 350 Program Therapy site P1 SHOULDER NECK BACK ELBOW HIP ANKLE FOOT WRIS...

Page 8: ...ime 2 sec ramp down time 2 sec rest time 4 sec Program 1 Program Work Time Pulse Rate Pulse Width Waveform Wave Description Therapy Site Min Hz s µ Shoulder 30 50 200 Synchronous Work time 13 sec ramp up 1 sec ramp down 2 sec Rest time 10 sec Neck 30 2 60 200 Simple Cycletime 12s span 100 Modulate percentage 50 Rate Pulse stayinthe2 10Hzrange for1 3ofthecycletime 4seconds astherate modulatesdownto...

Page 9: ...sec rest time 10 sec Foot 30 65 200 Synchronous Worktime10sec rampup time 2 sec ramp down 2 sec rest time 8 sec Wrist 30 50 350 Alternating Work time 6 sec ramp up time 1 sec ramp down sec 1 rest time 6 sec Knee 30 40 350 Synchronous Work time 11 sec ramp up time 4 sec ramp down time 2 sec Rest time 15 sec U1 Default Phase Pulse rate Pulse Width Work time Rest Time Hz s s s µ 1 25 30 3 5 1 0 220 2...

Page 10: ...et them air dry to help them last longer 3 Cleaning the electrode cords Clean the electrode cords by wiping them with damp cloth Coating them lightly with talcum powder will reduce tangles and help them last longer Reusable self adhesive electrodes 19 MASSAGE PROGRAMS PRESET ONLY CONTINUED Note Treatment length for all options below is 30 minutes CLEANING AND MAINTENANCE 1 Stimulator Wipe the stim...

Page 11: ...device and connect the cable Battery is low Charge battery 20 TROUBLESHOOTING If your device does not seem to be operating correctly refer to the chart below to determine what may be wrong If none of these measures correct the problem the device should be serviced Problem Possiblecause Solution Displays fail to light up Low Battery Battery may be low and needs to be charged Weak or no stimulation ...

Page 12: ... 5 C to 40 C with a relative humidity of 30 85 Atmospheric pressure from 700 hpa to 1060 hpa Storage conditions 10 C to 50 C with a relative humidity of 10 90 atmospheric pressure from 700 hpa to 1060 hpa Dimensions 117 x 60 x 25 mm without belt clip Weight including battery 107 g 23 ELECTRODE PLACEMENT GUIDE Muscle Group Placementof Electrodes Body Positionningand Voluntary Start of the Contracti...

Page 13: ...e muscular contraction phase dynamically to associate the treatment with sit ups In this case always work with your knees bent so as to not accentuate the lumbar region arch lordosis Strongly contract your abdominal muscles while vigorously trying to raise your head and shoulders from their support 25 1 2 1 2 1 2 ELECTRODE PLACEMENT GUIDE Muscle Group Placementof Electrodes Body Positionningand Vo...

Page 14: ... into the armrests 1 2 Biceps Sit with your forearms resting on armrests and the palms of your hands facing up Use a fixing system to avoid any movement or flexion of your elbows during stimulation Vigorously contract your biceps while trying strongly to move the palms of your hands towards your shoulders ELECTRODE PLACEMENT GUIDE Muscle Group Placementof Electrodes Body Positionningand Voluntary ...

Reviews: