WARNING
Patient safety risk. The EarlySense system is not intended for
high-risk patients in coronary or respiratory distress who require continuous
monitoring of heart function or CO2. For these patients, the most reliable
method of patient monitoring involves close personal surveillance and/or
equipment suited to that type of monitoring.
WARNING
Patient injury risk. The patient should not have direct contact
with the bed sensor. A mattress, mattress pad, or mattress cover should
always be placed as a barrier between the sensor and the patient. Check
patients regularly to ensure that direct contact with the sensor does not
occur.
WARNING
Patient safety risk. Provide careful oversight when the
EarlySense System is used with children.
CAUTION
Sensor malfunction risk. Do not use EarlySense with patients
who weigh more than 200 kg (440 pounds).
CAUTION
Sensor malfunction or inaccurate measurement risk. Patients
should not share a bed with another person during EarlySense monitoring.
CAUTION
Inaccurate measurement risk. The quality of measurements
might be affected by cable lengths. Do not shorten or extend the cable.
CAUTION
Sensor malfunction risk. Do not allow patients to jump on the
bed or the bed sensor.
CAUTION
Sensor malfunction risk. For patients with poor bladder function
or control, take care when placing the bed sensor under the mattress,
mattress pad, or mattress cover to prevent urine from contacting the
sensor.
CAUTION
Sensor malfunction risk. Do not create sharp bends in the cable
to prevent tearing or breaking the shielding.
1. Verify that you are using the Continuous Monitoring profile.
Note
If "Allow profile change" is enabled in Advanced settings, the
device attempts to shift automatically to the Continuous
Monitoring profile when you connect the sensor cable.
2. Touch the patient type control on the right side of the Patient frame.
The Patient Summary tab appears.
3. Select the patient type, then touch
OK
.
Note
Patient movement monitoring is not available for neonatal/
infant patients.
4. If sensor is not plugged into the device, align the EarlySense cable connector with
one of the EarlySense ports on the right side of the monitor. Insert the cable
connector until it clicks into place. Also check the stress relief connector on the cable
to ensure that both parts of the cable are tightly connected.
Directions for use
Patient monitoring 119
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......