Reset (pause or turn off) audio alarms
WARNING
Do not pause or turn off an audible alarm if patient safety
might be compromised.
Audio alarm characteristics
•
After you reset an audio alarm, some tones do not return, but others return after a
pause interval if the condition that caused the alarm persists.
•
If a new alarm condition occurs during a pause interval, a new audio tone occurs.
•
If an audio alarm is not paused or turned off after a period of time, a buzzer
accompanies the tone.
Pause or turn off an audio alarm
1. In the Device Status area, touch
.
•
Visual indications remain in the parameter frame until the condition is corrected,
the next measurement is taken, or the alarm is cancelled.
•
In the Device Status area, if the icon changes to
and the message
remains, the timer counts down a fixed pause interval of 60 seconds. The alarm
tone returns after the pause interval.
A longer pause interval can be configured in the Advanced tab. You start the
configured pause interval by touching
.
If you responded to an NIBP alarm and multiple NIBP limits have been
exceeded, the alarm tone pauses, the first message clears, and the next NIBP
limit message appears with a countdown timer. A new NIBP alarm tone sounds
after the countdown unless you touch
to dismiss each remaining NIBP
limit message.
2. If multiple alarms are active, a multiple alarm toggle appears in the Device Status
area. The monitor will either autoscroll through the alarm messages, displaying each
for approximately 4 seconds in order of priority, or you can move through the alarms
manually. Respond to multiple alarms manually as follows:
a. Touch
to pause all audio alarms.
The alarm icon changes to
and the timer counts down a fixed pause
interval of 60 seconds. The alarm tone returns after the pause interval.
b. Touch
in the Device Status area to view each alarm in the stack.
106 Alarms
Welch Allyn Connex
®
Devices
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......