EarlySense specifications
EH = 80% to 100%
Accuracy
Adult:
0 = 100%, L = 100%, M = 81%, H = 100%, EH = 96%
Pediatric:
0 = 100%, L = 100%, M = 81%, H = 86%, EH = 94%
Averaging period
15 seconds
Respiration rate
Unit of measure
breaths per minute
Measurement range
6 to 45 breaths per minute
1
Accuracy
2
±4% or ±1.5 breaths per minute, whichever is greater
Averaging period
1 minute
Pulse rate
Unit of measure
beats per minute
Measurement range
30 to 170 beats per minute
1
Accuracy
2
±4% or ±5 beats per minute, whichever is greater
Averaging period
1 minute
1
The system detects pulse rate that is > 1.8 times the respiration rate.
2
Total system accuracy including undetected signals equals 90%.
Alarm limits
Upper limit range of entry
Lower limit range of entry
Systolic
Adult: 30 to 258 mmHg (4.0 to 34.4 kPa). Factory
default: 220 mmHg (29.3 kPa).
Pediatric: 32 to 160 mmHg (4.3 to 21.3 kPa).
Factory default: 145 mmHg (19.3 kPa).
Neonate: 27 to 120 mmHg (3.6 to 16.0 kPa).
Factory default: 100 mmHg (13.3 kPa).
Adult: 28 to 256 mmHg (3.7 to 34.1 kPa). Factory
default: 75 mmHg (10.0 kPa).
Pediatric: 30 to 158 mmHg (4.0 to 21.1 kPa).
Factory default: 75 mmHg (10.0 kPa).
Neonate: 25 to 118 mmHg (3.3 to 15.7 kPa).
Factory default: 50 mmHg (6.7 kPa).
Diastolic
Adult: 22 to 235 mmHg (2.9 to 31.3 kPa). Factory
default: 110 mmHg (14.7 kPa).
Pediatric: 17 to 130 mmHg (2.3 to 17.3 kPa).
Factory default: 100 mmHg (13.3 kPa).
Neonate: 12 to 105 mmHg (1.6 to 14.0 kPa).
Factory default: 70 mmHg (9.3 kPa).
Adult: 20 to 233 mmHg (2.7 to 31.1 kPa). Factory
default: 35 mmHg (4.7 kPa).
Pediatric: 15 to 128 mmHg (2.0 to 17.1 kPa).
Factory default: 35 mmHg (4.7 kPa).
Neonate: 10 to 103 mmHg (1.3 to 13.7 kPa).
Factory default: 30 mmHg (4.0 kPa).
MAP
Adult: 22 to 255 mmHg (2.9 to 34.0 kPa). Factory
default: 120 mmHg (16.0 kPa).
Adult: 20 to 253 mmHg (2.7 to 33.7 kPa). Factory
default: 50 mmHg (6.7 kPa).
Directions for use
Specifications 233
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......