Change the battery (Connex VSM)
Before removing the battery, shut down the monitor.
1. Turn the monitor upside down to access the battery cover.
2. Locate the battery cover, indicated by
.
3. Insert a coin into the slot and push to open. Choose a coin that fits comfortably in
the slot.
4. Pull the battery out by pulling the battery label, which is visible when you open the
battery cover.
5. Slide in the new battery. Ensure that you insert the new battery in the same
orientation as the old battery.
6. Replace the battery cover by positioning the end below
and then pressing
firmly on the opposite end.
Directions for use
Maintenance and service 181