a. Touch the
Settings
tab.
b. Touch the
Advanced
tab.
c. Enter the
Advanced settings code
.
d. Touch
OK
.
The General tab appears.
2. Touch the
Data Management
tab.
3. Touch the
Clinician
tab.
4. Specify settings.
Setting
Action/Description
Label
Select a type of clinician identification label for display
on the Home tab:
Full name
,
Abbreviation
,
Clinician ID
, or
Symbol only
.
Require clinician ID to save readings
Make entering a clinician ID a prerequisite for saving
measurements. If they fail to enter identification, the
monitor prompts them when they try to save
measurements. Clinicians can enter clinician
identification on the Clinician tab.
Clear clinician information on manual save
Specify that the monitor clears the selected clinician
after a clinician manually saves measurements from the
Home tab. Clinician information clears from the Clinician
tab and the Device Status area.
Search by clinician ID
Enable the monitor to query the network for clinician
information based on ID. The monitor initiates the search
when the clinician enters or scans the ID from the
Clinician tab. Returned clinician information populates
the Device Status area and fields on the Clinician tab.
Select
Require password
to require clinicians to
enter their password, in addition to ID, on the Clinician
tab. The monitor uses the ID and password combination
to query the network for clinician information.
Select
Require clinician ID match to save
measurements
to require that the selected
clinican ID matches a clinician ID in an external host
system before you can save measurements.
Store clinician information for ___ hours
Select to enable a local cache of clinician login
information to speed up and simplify future clinicial
logins. Use the keypad to enter the number of hours to
store this clinician cache.
5. Do one of the following:
•
To continue in the Advanced Settings, touch another tab.
•
To exit the Advanced Settings and return to the Home tab, touch
Exit
.
Specify clinical data settings
1. Access the Advanced Settings.
a. Touch the
Settings
tab.
b. Touch the
Advanced
tab.
c. Enter the
Advanced settings code
.
d. Touch
OK
.
Directions for use
Advanced settings 201
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......