Audio
(vertical tab)
Allow user to turn off general audio
Select to allow clinicians to turn off all audio notification
for alarms. This control is on the Alarms tab (on the
General tab).
Minimum alarm volume
Select the minimum alarm volume available. If you select
High
, then
Medium
and
Low
are not available to
the clinician.
Audio pause time
Specify the amount of pause time that is added to the
60-second pause time. When a clinician pauses an audio
alarm tone, the tone is paused for the combined amount
of time.
Enable audio for low priority alarms
Select to enable the alarm to sound for low priority
alarms. If this setting is disabled, the audio for very low
priority alarms is automatically disabled.
Enable audio for very low priority alarms
Select to enable the alarm to sound for very low priority
alarms. This setting is available only when "Enable audio
for low priority alarms" is selected.
Allow user to turn on patient rest mode
Select to allow the user to turn off or turn on this option
on the Alarms tab. This option is only available in the
Continuous Monitoring profile and when the user has
enabled night mode.
Allow host to turn on patient rest mode
Select to allow the host to turn off or turn on this option
on the Alarms tab. This option is only available in the
Continuous Monitoring profile and when the device is
connected to the Central Station.
Delays
(vertical tab)
Note
Alarm delays are available only for the parameters
supported by your configuration.
SpO2 alarm condition delay
Specify the minimum amount of time that an SpO2 alarm
condition must be active before audio and visual signals
occur.
SatSeconds is available with Nellcor SpO2 sensors. If
you select Off, 10-, 15-, or 30-second delay, SatSeconds
is disabled, and it is removed from the SpO2 tab in the
Alarms tab.
SpO2 pulse rate alarm condition delay
Specify the minimum amount of time that a Pulse rate
alarm condition measured by the SpO2 sensor must be
active before audio and visual signals occur.
Motion pulse rate alarm condition delay
Specify the minimum amount of time that a Pulse rate
alarm condition measured by the patient movement
sensor must be active before audio and visual signals
occur.
SpHb alarm condition delay
Specify the minimum amount of time that an SpHb alarm
condition must be active before audio and visual signals
occur.
Motion respiration alarm condition delay
Specify the minimum amount of time that a respiration
alarm condition must be active before audio and visual
signals occur (EarlySense configuration).
Motion pulse rate low confidence alarm delay
Specify the minimum amount of time that a pulse rate
low confidence alarm condition must be active before
audio and visual signals occur (EarlySense
configuration).
Motion respiration low confidence alarm delay
Specify the minimum amount of time that a respiration
low confidence alarm condition must be active before
Directions for use
Advanced settings 189
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......