Message
Possible cause
Suggested action
Unable to detect new temperature.
Retry measurement.
The Braun thermometer is removed from the dock
or not seated properly in the dock
Return the thermometer to the
dock or adjust the thermometer
in the dock.
The Braun dock is disconnected from the device Connect the USB cable for the
Braun dock to the device.
Thermometer might be docked
improperly. Check contacts and
connections.
The Braun thermometer is removed from the dock
or not seated properly in the dock
Return the thermometer to the
dock or adjust the thermometer
in the dock.
The Braun dock is disconnected from the device Connect the USB cable for the
Braun dock to the device.
Weight scale messages
Message
Possible cause
Suggested action
Weight scale not functional. Call
for service.
The weight scale is not operating properly.
Call for service.
Physical assessment instrument handles
Symptom
Possible cause
Suggested action
The lamp does not illuminate
There is no lamp in the handle head
Install a lamp in the handle head.
The lamp is burned out
Install a new lamp.
The other handle is off the cradle
Place the other handle in the cradle.
The system is not powered up
Power up the system.
The platform handle controller PCBA is
defective
Call service.
The handle assembly is defective
Call service.
The lamp is too dim
The rheostat setting is too low
Increase the rheostat setting.
The platform handle controller PCBA is
defective
Call service.
The handle assembly is defective
Call service.
The lamp is too bright
The rheostat setting is too high
Decrease the rheostat setting.
The platform handle controller PCBA is
defective
Call service.
The handle assembly is defective
Call service.
214 Troubleshooting
Welch Allyn Connex
®
Devices
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......