5. (Optional) Change the name of the program:
a. Touch the keyboard icon.
b. Enter the name of the program and touch
OK
.
6. Specify the settings for this program.
Setting
Action/Description
Readings to average
Select readings to include in the average. You must
select at least two readings.
The last reading you select is the final reading of the
program. For example, if you select Readings 3, 4, and 5,
the program takes five readings.
Unselected readings—in this example, Readings 1 and 2
—are excluded ("discarded") from the average.
Delay to start
Enter the period between the start of the program (the
moment that the "Start intervals" button is selected) and
the start of the first reading.
Time between
Enter the period between the end of one reading and the
start of the next reading.
Keep if + or -
Enter the range that the program uses to establish the
baseline reading. For more information about how this
setting affects the program, refer to the section
"Excluded readings" below.
The settings you selected appear in the Summary area.
7. Do one of the following:
•
To continue in the Advanced Settings, touch another tab.
•
To exit the Advanced Settings and return to the Home tab, touch
Exit
.
The program is now available for use.
Specify advanced temperature settings
1. Access the Advanced Settings.
a. Touch the
Settings
tab.
b. Touch the
Advanced
tab.
c. Enter the
Advanced settings code
.
Directions for use
Advanced settings 197
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......