The General tab appears.
2. Touch the
Parameters
tab.
3. Touch the
RRa
tab.
4. Specify settings.
Setting
Action/Description
RRa averaging
Select the default averaging option that provides the
desired visibility of subtle variations in RRa
measurements.
Freshness timeout
Select the period of time the device should wait before
alarming while it attempts to obtain a valid RRa reading.
5. Do one of the following:
•
To continue in the Advanced Settings, touch another tab.
•
To exit the Advanced Settings and return to the Home tab, touch
Exit
.
Specify advanced SpO2 settings
1. Access the Advanced Settings.
a. Touch the
Settings
tab.
b. Touch the
Advanced
tab.
c. Enter the
Advanced settings code
.
d. Touch
OK
.
The General tab appears.
2. Touch the
Parameters
tab.
3. Touch the
SpO2
tab.
4. Specify settings.
Setting
Action/Description
Default view
Select a numeric view or a waveform view as the default
primary SpO2 display on the Home tab.
Default response
Select the default speed of response to changes in SpO2
measurements.
Sweep speed default
Select the default waveform sweep speed for the SpO2
display in the Home tab.
Allow low perfusion alarm
Select to enable Masimo low perfusion alarm.
5. Do one of the following:
•
To continue in the Advanced Settings, touch another tab.
•
To exit the Advanced tabs and return to the Home tab, touch
Exit
.
Specify advanced pulse rate settings
1. Access the Advanced Settings.
a. Touch the
Settings
tab.
b. Touch the
Advanced
tab.
c. Enter the
Advanced settings code
.
d. Touch
OK
.
Directions for use
Advanced settings 193
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......