CAUTION
Store the thermometer and probe covers in a dry location, free
from dust and contamination and away from direct sunlight. Keep the
ambient temperature at the storage location fairly constant and within the
range of 50°F to 104°F (10°C to 40°C).
Take a temperature at the ear site
WARNING
Probe covers are single-use only. Re-use of a probe cover may
result in spread of bacteria and cross-contamination.
WARNING
Inaccurate measurement risk. Use only Braun ThermoScan
probe covers with this thermometer.
WARNING
Inaccurate measurement risk. Frequently inspect the probe
window and keep it clean, dry, and undamaged. Fingerprints, cerumen,
dust, and other contaminants reduce the transparency of the window and
result in lower temperature measurements. To protect the window, always
keep the thermometer in the accessory dock when the thermometer is not
in use.
CAUTION
Inaccurate measurement risk. Before taking a temperature
measurement, make sure that the ear is free from obstructions and excess
cerumen build-up.
CAUTION
Inaccurate measurement risk. The following factors can affect
ear temperature measurements for up to 20 minutes:
•
The patient was lying on his or her ear.
•
The patient's ear was covered.
•
The patient was exposed to very hot or very cold temperatures.
•
The patient was swimming or bathing.
•
The patient was wearing a hearing aid or an ear plug.
CAUTION
Inaccurate measurement risk. If ear drops or other ear
medications have been placed in one ear canal, take the temperature in the
untreated ear.
Note
A temperature measurement taken in the right ear might differ from a
measurement taken in the left ear. Therefore, always take the temperature
in the same ear.
Note
When the monitor receives an ear temperature measurement, it displays
the measurement on the Home tab. If the Home tab already contains a
temperature measurement, the new measurement overwrites it.
To take a measurement and transfer it to the monitor:
1. Make sure that the monitor is powered on.
2. Remove the ear thermometer from the accessory dock.
The thermometer turns on automatically when you remove the thermometer from
the dock.
Note
If the thermometer has been undocked, it turns on after you
complete step 4.
3. Locate the probe cover box in the accessory dock.
160 Patient monitoring
Welch Allyn Connex
®
Devices
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......