audio and visual signals occur (EarlySense
configuration).
etCO2 alarm condition delay
Specify the minimum amount of time that an etCO2
alarm condition must be active before audio and visual
signals occur.
Respiration alarm condition delay
Specify the minimum amount of time that an RR alarm
condition must be active before audio and visual signals
occur.
No breath detected alarm delay
If configured with a CO2 module, specify the amount of
time for the device to wait after receiving a "valid
breath" message before activating the "No breath
detected" physiological alarm.
If configured for RRa, specify the time that the Masimo
module will wait before sending a respiration pause
event, which will result in a "No breath detected"
physiological alarm.
Adult no breath detected alarm delay
Specify the time that an adult no breath detected
condition must be active in an Oridion CO2 configuration
before audio and visual signals occur.
Pediatric no breath detected alarm delay
Specify the time that a pediatric no breath detected
condition must be active in an Oridion CO2 configuration
before audio and visual signals occur.
Neonate no breath detected alarm delay
Specify the time that a neonate no breath detected
condition must be active in an Oridion CO2 configuration
before audio and visual signals occur.
4. Do one of the following:
•
To continue in the Advanced Settings, touch another tab.
•
To exit the Advanced Settings and return to the Home tab, touch
Exit
.
Specify advanced display settings
1. Access the Advanced Settings.
a. Touch the
Settings
tab.
b. Touch the
Advanced
tab.
c. Enter the
Advanced settings code
.
d. Touch
OK
.
The General tab appears.
2. Touch the
Display
tab.
3. Specify settings.
Setting
Action/Description
Display lock
Specify the required period of clinician inactivity before
the touchscreen locks.
Display power saver
Specify the required period of monitor inactivity before
the display turns off.
Clinician interactions, new vital sign measurements, or
alarm conditions automatically turn on the display.
Device power down
Specify the required period of monitor inactivity before
the monitor turns off.
4. Do one of the following:
190 Advanced settings
Welch Allyn Connex
®
Devices
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......