Message
Possible cause
Suggested action
Remove the sensor from the
patient and reapply.
SpO2 sensor expires in….
Note
This message only appears
on devices configured with SpHb.
The SpO2 sensor will expire soon
Replace the SpO2 sensor.
Temperature messages
Message
Possible cause
Suggested action
Connect temperature probe.
No probe is connected
Connect a temperature probe and
retry.
The probe is faulty
Replace the temperature probe.
The temperature module returned a connect
probe message
Connect a temperature probe and
retry. If a probe is already
connected, replace the probe.
Insert correct color-coded probe
well.
The probe well is missing
Insert a temperature probe well.
Replace temperature probe.
The probe is faulty
Replace the temperature probe.
Temperature not functional. Call
for service.
A module error occurred
Call for service.
Temperature time limit exceeded.
Retry temperature measurement.
The direct mode timed out
Remove the probe from the
measurement site.
Tissue contact lost.
The probe has lost contact with the patient's
tissue
Touch
OK
to dismiss the
message. Return the probe to the
probe well and retry patient
temperature measurement.
Ensure that the probe makes
proper contact with the patient's
tissue.
Retry temperature measurement.
Note
This message often
accompanies other temperature
messages.
A probe heater or data error occurred
Retry the temperature
measurement. If the problem
persists, replace the probe.
User settings require adjustment
Adjust the user settings and
retry.
The ambient temperature is out of range
Operate the monitor within the
specified temperature range.
Retry patient temperature
measurement.
The SureTemp temperature module is faulty
Call for service.
Directions for use
Troubleshooting 213
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......