Note
The multiple alarm toggle button displays the number of
active alarms inside the alarm icon. A set of dots
indicating the display order of alarms from highest (left)
to lowest (right) priority (as well as the most recent in
the case of multiple alarms of the same priority) appears
below it.
Cancel a paused alarm
You can clear a paused alarm from the Home tab. If the condition that caused the alarm
persists, a new alarm appears with visual and audio notification.
1. On the Home tab, touch the alarm limits control in the selected parameter frame.
The Alarms tab for that parameter appears.
2. Touch
.
The alarm clears.
3. Touch
.
4. Touch the Home tab and confirm that
appears in the parameter frame.
Adjust vital sign alarm limits
You can adjust vital sign alarm limits or turn off alarm limit checking for individual
parameters.
WARNING
Alarm limits are user-adjustable. For alarms to function
properly, you must set or verify alarm limits appropriate for each patient,
taking into account the patient's condition and acute care needs. Each time
the monitor is powered on, you must check that the alarm settings are
appropriate for your patient before you start monitoring.
CAUTION
Loss of power will cause the monitor to return to default
settings. Each time you power up the monitor, you must set alarm limits
appropriate for your patient.
1. On the Home tab, touch the alarm limits control in the selected parameter frame. For
example, to adjust the NIBP alarm limits, touch
.
2. Adjust vital sign alarm limits.
•
To adjust a limit: Enter the desired upper and lower alarm limits using the up/
down arrow keys or the keypad.
Directions for use
Alarms 107
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......