4. Firmly push the probe tip into the probe cover box.
When the probe cover is in place, previously undocked thermometers turn on
automatically.
5. Proceed as follows based on your Braun thermometer model:
•
Braun 4000: Wait for the ready beep and three dashes to appear on the
thermometer display.
•
Braun 6000: Wait for the ready beep, three dashes to appear on the
thermometer display, and the ring around the Measure button to turn green.
6. Fit the probe snugly into the ear canal and proceed as follows based on your Braun
thermometer model:
•
Braun 4000: Push and release the
Start
button.
•
Braun 6000: Push and release the
Measure
button.
•
If the probe is positioned correctly in the ear canal the ExacTemp light flashes.
When the thermometer detects an accurate measurement, the ExacTemp light
is continuously on, a long beep signals the end of the measurement, and the
display shows the result.
•
If the probe is positioned incorrectly in the ear canal or is moved during the
measuring process, the ExacTemp light goes out, a sequence of short beeps
sounds, and the error message POS (position error) appears.
7. When you are finished taking the temperature, press the ejector button to eject the
used probe cover.
8. Return the thermometer to the accessory dock.
After the transfer is complete, the temperature and the temperature scale appear on
the Home tab according to the monitor settings.
Note
Only the latest measurement is transferred to the monitor.
Note
Measurements that have already been transferred to the
monitor cannot be transferred again.
For more information about thermometer functionality, refer to the thermometer
manufacturer's directions for use.
Change the temperature scale on the ear thermometer
To switch from Celsius to Fahrenheit, refer to the thermometer manufacturer's
directions for use.
Charge the ear thermometer battery
To charge the battery pack:
•
Place the thermometer in the accessory dock.
•
Make sure that the monitor is connected to AC power.
•
Make sure that the monitor is powered on.
The LED on the dock indicates the charging status of the battery pack.
PRO 6000 model
PRO 4000 model
•
Amber: The battery pack is charging.
•
Green: The battery pack is charged.
Directions for use
Patient monitoring 161
Summary of Contents for Connex IWS
Page 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Page 10: ...4 Introduction Welch Allyn Connex Devices...
Page 14: ...8 Symbols Welch Allyn Connex Devices...
Page 20: ...14 Screen elements Welch Allyn Connex Devices...
Page 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Page 72: ...66 Navigation Welch Allyn Connex Devices...
Page 96: ...90 Profiles Welch Allyn Connex Devices...
Page 106: ...100 Patient data management Welch Allyn Connex Devices...
Page 120: ...114 Alarms Welch Allyn Connex Devices...
Page 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Page 212: ...206 Advanced settings Welch Allyn Connex Devices...
Page 244: ...238 Specifications Welch Allyn Connex Devices...
Page 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Page 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Page 266: ......