Remote Control
R&S
®
FPC
292
User Manual 1178.4130.02 ─ 07
Suffix:
<n>
.
irrelevant
<k>
Return values:
<Status>
•
0 for PASS
•
1 for FAIL
Example:
//Perform a measurement and query the result of the check for
limit line 1.
CALC:LIM:UPP:THR -70
INIT;*WAI
CALC:LIM1:FAIL?
Usage:
Query only
Manual operation:
See
See
See
See
See
See
CALCulate<n>:LIMit<k>:LOWer:SELect
<LimitLine>
This command selects the lower limit line.
You can either select a limit line that is already stored on the internal memory or create
a new one with
.
Suffix:
<n>
.
irrelevant
<k>
Setting parameters:
<LimitLine>
String containing the file name of the lower limit line.
Example:
//Select the lower limit line
FREQ:SPAN 0HZ
CALC:LIM:LOW:SEL 'GSM_Lower.rellim'
Usage:
Setting only
Manual operation:
See
See
See
CALCulate<n>:LIMit<k>:LOWer:THReshold
<Level>
This command defines the level of a lower threshold limit line.
Suffix:
<n>
.
irrelevant
Remote Commands of the Spectrum Application