Remote Control
R&S
®
FPC
316
User Manual 1178.4130.02 ─ 07
Example:
//Query frequency deviation
CALC:MARK:FUNC:DDEM:MDEV?
Usage:
Query only
CALCulate<n>:MARKer:FUNCtion:DDEMod:MERRor[:RESult]?
This command queries the magnitude error of the measured signal.
Prerequisites for this command
●
).
Suffix:
<n>
.
irrelevant
Return values:
<MagnitudeError>
<numeric value>
Default unit: PCT
Example:
//Query magnitude error
CALC:MARK:FUNC:DDEM:MERR?
Usage:
Query only
CALCulate<n>:MARKer:FUNCtion:DDEMod:MINDex[:RESult]?
This command queries the modulation index of the measured signal.
Prerequisites for this command
●
Select ASK demodulation (
Suffix:
<n>
.
irrelevant
Return values:
<ModIndex>
<numeric value>
Default unit: PCT
Example:
//Query modulation index
CALC:MARK:FUNC:DDEM:MIND?
Usage:
Query only
CALCulate<n>:MARKer:FUNCtion:DDEMod:MOERror[:RESult]?
This command queries the modulation error.
Suffix:
<n>
.
irrelevant
Return values:
<ModError>
<numeric value>
Default unit: PCT
Remote Commands of the Digital Demodulation