![Rohde & Schwarz FPC Series User Manual Download Page 321](http://html.mh-extra.com/html/rohde-and-schwarz/fpc-series/fpc-series_user-manual_1477824321.webp)
Remote Control
R&S
®
FPC
318
User Manual 1178.4130.02 ─ 07
TRACe[:DATA]?
This command queries the trace data (measurement results in graphical result dis-
plays).
Return values:
<Results>
Returns one value for each measurement point.
The type of value depends on the measurement and result dis-
play.
"Symbols": returns the symbol stream.
"Modulation Deviation": returns one value (frequency deviation)
for each symbol in Hz.
"Modulation Depth": returns one value (modulation depth) for
each symbol in PCT.
"Modulation Error": returns one value for each symbol. The unit
is either Hz (FSK) or PCT (ASK).
Example:
//Query results
TRAC?
Usage:
Query only
19.15.2
Measurement Configuration
................................................................................... 318
.....................................................................................318
......................................................................................319
.................................................................................... 319
............................................................................................... 320
[SENSe:]DDEMod:BURSt[:STATe]
<State>
This command turns burst processing on and off.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
OFF
Example:
//Turn on burst processing
DDEM:BURS ON
Manual operation:
See
[SENSe:]DDEMod:FILTer:ALPHa
<Alpha>
This command defines the roll-off factor of the measurement filter (alpha or BT value).
Remote Commands of the Digital Demodulation