The Receiver Application
R&S
®
FPC
135
User Manual 1178.4130.02 ─ 07
The bargraph measurement measures the signal level of a
and shows
the results in a bargraph.
The bargraph result display is a basic result display that shows the signal level at a
specific frequency. The signal level is shown both as a numeric value and graphically
as a bargraph.
The results in the bargraph are shown as soon as you enter the receiver application
(for the frequency that is currently selected). The results are updated continuously. The
current signal level is represented by light blue vertical lines in the bargraph. The unit
of the displayed power level is variable and depends on the unit you have set.
The scale of the bargraph depends on the current reference level and display range.
In addition to the current signal level, the bargraph also shows the highest signal level
that has been measured since the measurement has begun (
remains effective, even if you change any receiver settings (for example the frequency
or the attenuation) and is only reset when you deliberately do so.
The max hold information is represented by a dark blue vertical line in the bargraph.
The max hold information is also displayed in a dark blue font above the bargraph.
Figure 15-1: Diagram for the bargraph measurement, including maxhold information
Selecting bargraph measurements
1. Press the "Meas" key to open the measurement menu.
2. Select the "Bargraph" menu item.
The R&S
FPC shows the signal level measured at the current receiver frequency.
Displaying a second bargraph
You can display a second bargraph with a different configuration (the detector, for
example).
1. Press the "Meas" key to open the measurement menu.
2. Select the "Show Bargraph 2" menu item.
Measurements and Result Displays