Remote Control
R&S
®
FPC
226
User Manual 1178.4130.02 ─ 07
Parameters:
<FileName>
String containing the path and name of the dataset.
The path and file name comply with Windows conventions. Note
that the folder and filename references in the string are case-
sensitive.
Example:
//Remove write-protection from a file
SYST:SET:UNL '\Public\funny\result.set'
19.9
Signal Source Configuartion
................................................................................................226
......................................................................................227
SOURce:TG:AUTO
<State>
This command turns the tracking generator on and off.
Prerequisites for this command
●
).
●
Select a span > 0.
Using the tracking generator is not possible for time domain measurements (span
= 0). For time domain measurements, you have to turn off the tracking generator
and either generate a signal whose frequency is fix or is coupled to the center fre-
quency.
–
–
Coupled to center frequency:
Parameters:
<State>
ON | 1
Turns on the tracking generator. The R&S
FPC generates a sig-
nal whose frequency depends on the current measurement fre-
quency (signal source mode "Tracking Generator").
OFF | 0
Turns off the tracking generator. You can either generate a sig-
nal whose frequency is fix or is coupled to the center frequency.
Example:
//Turn on tracking generator
SOUR:TG ON
SOUR:TG:AUTO OFF
Manual operation:
See
SOURce:TG:FREQuency
<Frequency>
This command defines the frequency of the signal generated by the signal source.
Signal Source Configuartion