Remote Control
R&S
®
FPC
352
User Manual 1178.4130.02 ─ 07
Usage:
Event
STATus:QUEue[:NEXT]?
This command returns the earliest entry of the error queue and deletes it.
Positive error numbers indicate device-specific errors, negative error numbers are error
messages defined by SCPI. If the error queue is empty, the error number 0, "No error",
is returned.
This command is identical to the command
Return values:
<Error>
Example:
//Query error queue
STAT:QUE?
Usage:
Query only
STATus:OPERation[:EVENt]?
This command reads out the EVENt section of the OPERation register.
The command at the same time deletes the contents of the EVENt section.
Return values:
<Event>
Usage:
Query only
STATus:OPERation:ENABle
<SumBit>
This command controls the ENABle part of the OPERation register.
The ENABle part allows true conditions in the EVENt part of the status register to be
reported in the summary bit. If a bit is 1 in the enable register and its associated event
bit transitions to true, a positive transition will occur in the summary bit reported to the
next higher level.
Parameters:
<SumBit>
<numeric value> (integer)
Range:
0 to 65535
STATus:OPERation:NTRansition
<SumBit>
This command controls the Negative TRansition part of the OPERation register.
Setting a bit causes a 1 to 0 transition in the corresponding bit of the associated regis-
ter. The transition also writes a 1 into the associated bit of the corresponding EVENt
register.
Status Reporting System