![Rohde & Schwarz FPC Series User Manual Download Page 253](http://html.mh-extra.com/html/rohde-and-schwarz/fpc-series/fpc-series_user-manual_1477824253.webp)
Remote Control
R&S
®
FPC
250
User Manual 1178.4130.02 ─ 07
●
●
Parameters:
<OperationMode>
CW
Select the time domain (span = 0)
FIXed
Select the time domain (span = 0)
SWEep
Select the frequency domain (span > 0).
*RST:
SWEep
Example:
//Activate frequency domain measurements
FREQ:MODE SWE
Manual operation:
See
[SENSe:]FREQuency:OFFSet
<FrequencyOffset>
This command defines a frequency offset.
Parameters:
<FrequencyOffset>
<numeric value>
Range:
depends on the start frequency to 100 GHz
*RST:
0 Hz
Default unit: Hz
Example:
//Define a frequency offset
FREQ:OFFS 100MHZ
Manual operation:
See
See
[SENSe:]FREQuency:SPAN
<Span>
This command defines the frequency span.
If you define a span of 0 Hz, the R&S
FPC starts a measurement in the time domain.
Parameters:
<Span>
<numeric value>
Refer to the datasheet for details on the frequency range of your
R&S
FPC.
*RST:
fmax with fmax = maximum frequency
Default unit: Hz
Example:
//Define frequency span
FREQ:SPAN 10MHZ
Manual operation:
See
See
See
Remote Commands of the Spectrum Application