![Rohde & Schwarz FPC Series User Manual Download Page 266](http://html.mh-extra.com/html/rohde-and-schwarz/fpc-series/fpc-series_user-manual_1477824266.webp)
Remote Control
R&S
®
FPC
263
User Manual 1178.4130.02 ─ 07
Manual operation:
See
See
See
See
See
See
See
See
See
See
See
See
INITiate[:IMMediate]
The command initiates a new measurement sequence.
Effects of this command
●
In single sweep mode, with an average count > 0, the command restarts the mea-
surement.
●
When you have selected the maxhold, minhold or average trace mode the com-
mand resets the trace.
in single sweep mode, you can synchronize to the end of the measurement with *OPC,
*OPC? or *WAI. In continuous sweep mode, synchronization to the end of the mea-
surement is not possible. Thus, it is not recommended that you use continuous sweep
mode in remote control, as results like trace data or markers are only valid after a sin-
gle sweep end synchronization.
Example:
//Start a single sweep and wait to end of sweep
INIT:CONT OFF
INIT;*WAI
Usage:
Event
Manual operation:
See
See
See
See
See
See
See
See
See
See
See
See
Remote Commands of the Spectrum Application