![Rohde & Schwarz FPC Series User Manual Download Page 323](http://html.mh-extra.com/html/rohde-and-schwarz/fpc-series/fpc-series_user-manual_1477824323.webp)
Remote Control
R&S
®
FPC
320
User Manual 1178.4130.02 ─ 07
[SENSe:]DDEMod:SRATe
<SymbolRate>
This command defines the symbol rate.
Parameters:
<SymbolRate>
<numeric value> (integer only)
Default unit: Hz
Example:
//Define symbol rate
DDEM:SRAT 10KHZ
Manual operation:
See
[SENSe:]DDEMod:TIME
<Symbols>
This command defines the number of symbols to be evaluated.
Parameters:
<Symbols>
<numeric value> (integer only)
Range:
20 to depends on measurement filter
*RST:
100
Example:
//Define the number of symbols
DDEM:TIME 150
Manual operation:
See
19.15.3
Frequency
Commands to configure the frequency described elsewhere.
●
Center frequency:
●
Frequency stepsize:
●
Frequency offset:
19.15.4
Amplitude
Commands to configure amplitude described elsewhere.
●
Reference level:
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]:RLEVel
●
Preamplifier:
●
Attenuation:
INPut:ATTenuation
INPut:ATTenuation:AUTO
INPut:ATTenuation:MODE
Remote Commands of the Digital Demodulation