![Rohde & Schwarz FPC Series User Manual Download Page 237](http://html.mh-extra.com/html/rohde-and-schwarz/fpc-series/fpc-series_user-manual_1477824237.webp)
Remote Control
R&S
®
FPC
234
User Manual 1178.4130.02 ─ 07
Setting parameters:
<Standard>
String containing the file name of the standard.
Make sure to use the correct file extension (for exam-
ple
.chpstd
for channel power measurements or
.obwstd
for
occupied bandwidth measurements.
Example:
//Select the 3GPP WCDMA standard for channel power mea-
surements
CALC:MARK:FUNC:POW:PRES '3GPP WCDMA.chpstd'
Usage:
Setting only
Manual operation:
See
See
See
CALCulate<n>:MARKer<m>:FUNCtion:POWer:PRESet:CHECk?
This command checks if the current power measurement settings comply to the set-
tings defined in the currently selected standard.
Prerequisites for this command
●
Select a telecommunications standard for power measurements (
MARKer<m>:FUNCtion:POWer:PRESet
Note that the command only checks the parameters that are defined by the standard.
Suffix:
<n>, <m>
.
irrelevant
Return values:
<State>
0
R&S
FPC settings violate a standard configuration.
1
R&S
FPC settings comply with the standard.
Example:
//Query compliance to the standard currently in use
CALC:MARK:FUNC:POW:PRES '3GPP WCDMA.chpstd'
CALC:MARK:FUNC:POW:PRES:CHEC?
Usage:
Query only
CALCulate<n>:MARKer<m>:FUNCtion:POWer:RESult?
<Measurement>
This command queries the results of power measurements.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end before reading out the result. This is only possible in single sweep mode.
Prerequisites for this command
●
Select a power measurement (
CALCulate<n>:MARKer<m>:FUNCtion:POWer:
Remote Commands of the Spectrum Application