![Rohde & Schwarz FPC Series User Manual Download Page 331](http://html.mh-extra.com/html/rohde-and-schwarz/fpc-series/fpc-series_user-manual_1477824331.webp)
Remote Control
R&S
®
FPC
328
User Manual 1178.4130.02 ─ 07
Example:
//Activate DTF list
MEAS:MODE DTF
CALC:TRAN:DTF:PEAK ON
Manual operation:
See
CALCulate<n>:TRANsform:DTFault:PEAK:THReshold
<Threshold>
This command defines a threshold for the DTF list. Only peaks above that threshold
will appear in the DTF list.
Prerequisites for this command
●
Select the DTF measurement (
Suffix:
<n>
.
irrelevant
Parameters:
<Threshold>
<numeric value>
The unit depends on the measurement format (dB for return
loss, no unit for SWR).
*RST:
0
Example:
//Define threshold for DTF list
MEAS:MODE DTF
CALC:TRAN:DTF:PEAK:THR -10
Manual operation:
See
19.16.2
Amplitude
Commands to configure amplitude described elsewhere.
●
Auto Range:
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]:ADJust
........................................................................................ 329
DISPlay<n>:LOSS:REFerence:POSition
.......................................................................... 329
........................................................................................... 329
DISPlay<n>:MAGNitude:REFerence
................................................................................ 330
DISPlay<n>:MAGNitude:REFerence:POSition
.................................................................. 330
................................................................................... 331
DISPlay<n>:MAGNitude:Y:SPACing
.................................................................................331
...................................................................................... 332
DISPlay<n>:PHASe:REFerence:POSition
........................................................................ 332
.........................................................................................333
......................................................................................... 333
.......................................................................................... 334
DISPlay<n>:VSWR:Y:SCALe:MAXimum
.......................................................................... 334
DISPlay<n>:VSWR:Y:SCALe:MINimum
............................................................................335
Remote Commands of the VNA Application