Remote Control
R&S
®
FPC
331
User Manual 1178.4130.02 ─ 07
Suffix:
<n>
.
irrelevant
Parameters:
<RefPosition>
<numeric value> (integer only)
*RST:
8
Example:
//Define 5th grid line as the reference position for return loss
MEAS:MODE REFL
MEAS:FORM MAGN
DISP:MAGN:REF:POS 5
Manual operation:
See
DISPlay<n>:MAGNitude:Y:SCALe
<DisplayRange>
This command defines the range of the y-axis for the return loss measurement format.
Prerequisites for this command
●
Select reflection, transmission or DTF measurement (
●
).
●
Select a logarithmic scale (
DISPlay<n>:MAGNitude:Y:SPACing
(For a linear scale, the range is always 100
%.)
Suffix:
<n>
.
irrelevant
Parameters:
<DisplayRange>
<numeric value>
*RST:
100
Default unit: dB
Example:
//Define display range for return loss
MEAS:MODE REFL
MEAS:FORM MAGN
DISP:MAGN:Y:SPAC LOG
DISP:MAGN:Y:SCAL 50
Manual operation:
See
DISPlay<n>:MAGNitude:Y:SPACing
<Scaling>
This command selects scaling of the y-axis for the return loss measurement format.
Prerequisites for this command
●
Select reflection, transmission or DTF measurement (
●
).
Suffix:
<n>
.
irrelevant
Remote Commands of the VNA Application