Digital Demodulation
R&S
®
FPC
169
User Manual 1178.4130.02 ─ 07
Single Measurement
See
"Selecting the measurement mode"
Remote command:
Measurement type:
New measurement sequence:
17.6
Trigger Configuration
Access
: "Sweep" > "Trigger"
Remote commands to trigger measurements:
●
Trigger settings in the digital demodulation described elsewhere:
●
Chapter 14.7, "Trigger Configuration"
●
"External Rise":
Chapter 14.7, "Trigger Configuration"
●
"External Fall":
Chapter 14.7, "Trigger Configuration"
In addition to the trigger settings available in the spectrum application, the digital
demodulation application provides an I/Q power trigger source.
Using the I/Q power trigger
The I/Q power trigger initiates a measurement when the magnitude of the sampled I/Q
data exceeds a certain level.
1. Press the "Sweep" key to open the sweep menu.
2. Select the "Trigger" menu item to open the trigger menu.
3. Select the "IQ Power <x>" menu item.
The R&S
FPC opens an input field to define the trigger level.
4. Define an appropriate trigger level.
To make sure that the trigger initiates a measurement, we recommend a trigger
level that is about 6
dB lower than the actual expected signal level.
Note that a burst is only recognized if it is not too short and if the gap between two
bursts is not too short.
Trigger Configuration