![Rohde & Schwarz FPC Series User Manual Download Page 79](http://html.mh-extra.com/html/rohde-and-schwarz/fpc-series/fpc-series_user-manual_1477824079.webp)
The Spectrum Application
R&S
®
FPC
76
User Manual 1178.4130.02 ─ 07
The TDMA power measurement is a time domain measurement that determines the
power of one of the time slots or the power of a burst.
Selecting the TDMA power measurement
1. Press the "Meas" key to open the measurement menu.
2. Select the "Measurement Mode" menu item to select the measurement.
3. Select the "TDMA Power" menu item.
The R&S
FPC starts to measure the time domain power.
When you start a TDMA measurement, the R&S
FPC automatically reduces the span
to 0
Hz.
To prevent incorrect power measurements in the time domain, make sure that the
whole signal is within the selected resolution bandwidth. If the resolution bandwidth is
too small, the displayed power will be lower than the actual power.
The TDMA power measurement shows the results and various other information about
the measurement in the diagram header.
●
Power
Shows the power of the measured burst in dBm.
●
Burst Length
Shows the length of the measured burst in seconds.
Selecting a standard
A quick way to configure a measurement is to store a certain measurement configura-
tion in a file and later restore it. This way is especially useful, for example, if you test a
signal against the test scenarios of a telecommunications standard.
Configurations for common telecommunications standards are already provided with
the R&S
FPC firmware.
If you want to measure a different standard, you can also create custom configuration
files with the R&S
InstrumentView software and transfer them to the R&S
FPC. For
more information, refer to the documentation of the R&S
InstrumentView.
1. Press the "Meas" key to open the measurement menu.
2. Select the "Standard" menu item.
3. Select one of the available standards from the dialog box.
The R&S
FPC loads the configuration of the selected standard.
Depending on the selected file, the R&S
FPC automatically selects the span, reso-
lution bandwidth, video bandwidth, sweep time and detector required by the stan-
dard.
If you change the configuration, and one of the parameters differs from the configu-
ration in the active file, the R&S
FPC shows a blue dot in front of the standard dis-
play.
Measurements and Result Displays