![Rohde & Schwarz FPC Series User Manual Download Page 208](http://html.mh-extra.com/html/rohde-and-schwarz/fpc-series/fpc-series_user-manual_1477824208.webp)
Remote Control
R&S
®
FPC
205
User Manual 1178.4130.02 ─ 07
Parameters:
<Contents>
Contents of the service request enable register in decimal form.
Bit 6 (MSS mask bit) is always 0.
Range:
0 to 255
*STB?
Status byte query
Reads the contents of the status byte in decimal form.
Usage:
Query only
*TRG
Trigger
Triggers all actions waiting for a trigger event. In particular, *TRG generates a manual
trigger signal. This common command complements the commands of the
TRIGger
subsystem.
*TRG
corresponds to the
INITiate:IMMediate
command.
Usage:
Event
*TST?
Self-test query
Initiates self-tests of the instrument and returns an error code.
Return values:
<ErrorCode>
integer > 0 (in decimal format)
An error occurred.
0
No errors occurred.
Usage:
Query only
*WAI
Wait to continue
Prevents servicing of the subsequent commands until all preceding commands have
been executed and all signals have settled (see also command synchronization and
).
Usage:
Event
Common Commands