![Rohde & Schwarz FPC Series User Manual Download Page 185](http://html.mh-extra.com/html/rohde-and-schwarz/fpc-series/fpc-series_user-manual_1477824185.webp)
Vector Network Analyzer
R&S
®
FPC
182
User Manual 1178.4130.02 ─ 07
Defining the display range
The display range defines the scale of the vertical axis and therefore the amplitude
between two horizontal grid lines.
The unit depends on the measurement format.
When you change the display range, you can increase or decrease the amplitude the
R&S
FPC displays and, for example, include signal parts that are outside the displayed
screen area. The positions of the reference value and the trace do not change.
1. Press the "Ampt" key to open the amplitude menu.
2. Select the "Range/Ref Pos" menu item.
3. Select one of the menu items to select the display range you want.
● "Range": Defines the scope of the y-axis in a logarithmic scale (dB). By default,
the y-axis covers 100
dB, but you can increase or decrease the level range.
● "Linear": Changes the y-axis to a linear scale in %. The range is always 100
%.
● "SWR Scale": Defines the scope of the y-axis for the
in terms of a minimum and maximum value that is displayed on the y-axis.
● "Auto Range": Scales the y-axis in a way that the results fit ideally on the dis-
play. The R&S
FPC does this by determining the minimum and maximum trace
values and scaling the vertical axis according to these values.
4. Select the "Full Scale" menu item to restore the default display range.
18.4
Bandwidth Configuration
Access
: "BW"
Remote commands to configure the bandwidth:
●
The measurement bandwidth determines the noise figure of the receiver. Small band-
widths result in higher measurement dynamics for VNA measurements. However,
small bandwidths cause longer measurement times because of the settling time of the
filter.
Defining the measurement bandwidth
1. Press the "BW" key to open the bandwidth menu.
2. Select one of the "RBW: <x>" menu items.
● "RBW: Manual"
Opens an input field to select one of the available bandwidths.
If you enter a bandwidth that the R&S
FPC does not support, the R&S
FPC
rounds to the next available bandwidth. Refer to the data sheet for more infor-
mation on available resolution bandwidths
A manually selected RBW is indicated by a blue dot in front of the "RBW" label
in the status bar.
Bandwidth Configuration