Remote Control
R&S
®
FPC
229
User Manual 1178.4130.02 ─ 07
Example:
//Abort calibration
CAL:STAR? S11C
'Connect Open to Port 1'
CAL:ABOR
Usage:
Event
Manual operation:
See
CALibration<n>:CONTinue?
This command resumes calibration after a calibration standard has been connected.
Suffix:
<n>
.
irrelevant
Return values:
<Instructions>
String containing the instructions for the next calibration step.
You can find a list possible instructions in the table below.
Example:
//Resume calibration
CAL:STAR? S11C
'Connect Open to Port 1'
//Connect Open
CAL:CONT?
'Connect Short to Port 1'
Usage:
Query only
Manual operation:
See
Table 19-6: Return values: Possible instructions for calibration
'Connect Open to Port 1'
Connect the open to port 1 before resuming calibration.
'Connect Short to Port 1'
Connect the short to port 1 before resuming calibration.
'Connect Load to Port 1'
Connect the load to port 1 before resuming calibration.
'Connect Through'
Connect a through to ports 1 and 2.
'Disconnect DUT from port 1'
'Calibration Done'
Calibration has been finished. No further actions required.
CALibration<n>:MODE?
This command queries the current calibration type.
Suffix:
<n>
.
irrelevant
Return values:
<State>
0
Not calibrated or default calibration.
1
User calibration.
Calibration