Remote Control
R&S
®
FPC
251
User Manual 1178.4130.02 ─ 07
[SENSe:]FREQuency:SPAN:FULL
This command restores the full span.
Refer to the datasheet for details on the frequency range of your R&S
FPC.
Example:
//Restore full span
FREQ:SPAN:FULL
Usage:
Event
Manual operation:
See
See
[SENSe:]FREQuency:STARt
<Frequency>
This command defines the start frequency for measurements in the frequency domain
(span > 0).
Parameters:
<Frequency>
<numeric value>
Refer to the datasheet for details on the frequency range of your
R&S
FPC.
*RST:
0 Hz
Default unit: Hz
Example:
//Define start frequency
FREQ:STAR 20MHZ
Manual operation:
See
See
[SENSe:]FREQuency:STOP
<Frequency>
This command defines the stop frequency for measurements in the frequency domain
(span > 0).
Parameters:
<Frequency>
<numeric value>
Refer to the datasheet for details on the frequency range of your
R&S
FPC.
*RST:
fmax
Default unit: Hz
Example:
//Define stop frequency
FREQ:STOP 500MHZ
Manual operation:
See
See
Remote Commands of the Spectrum Application