Remote Control
R&S
®
FPC
257
User Manual 1178.4130.02 ─ 07
Parameters:
<Impedance>
50 | 75
*RST:
50
Default unit:
Ω
Example:
//Select input impedance
INP:IMP:PAD RAZ
INP:IMP 75
Manual operation:
See
See
See
INPut:IMPedance:PAD
<PadType>
This command selects a matching pad required for a 75
Ω
input impedance.
Parameters:
<PadType>
RAM | RAZ | HZTE
HZTE
75
Ω
FSH-Z38
RAM
75
Ω
RAM
RAZ
75
Ω
RAZ
Example:
//Select 75 Ω input impedance and the R&S RAZ as the match-
ing pad.
INPut:IMP 75;PAD RAZ
Manual operation:
See
See
See
[SENSe:]CORRection:TRANsducer<td>:SELect
<TransducerFactor>
This command selects a transducer factor.
Suffix:
<td>
.
Parameters:
<TransducerFactor> String containing the file name of the transducer factor. If the file
does not exist, the R&S
FPC creates a new transducer factor.
Example:
//Select primary transducer
CORR:TRAN1:SEL 'HE200A-20-200.pritrd'
//Select secondary transducer
CORR:TRAN2:SEL 'PreAmp.sectrd'
Manual operation:
See
See
Remote Commands of the Spectrum Application