CALCulate<n>:MARKer<m>:X
<Position>
This command positions a marker on a certain coordinate on the x-axis.
If one or more delta markers (2 to 6) are already active, the command turns these delta
markers into normal markers.
Note that it is possible to place the marker outside the visible trace. In that case, this
value is invalid.
If necessary, the corresponding delta marker is activated first.
Suffix:
<n>
.
irrelevant
<m>
Parameters:
<Position>
<numeric value>
The unit depends on the measurement and marker input mode.
Hz for measurements in the frequency domain.
s for measurements in the time domain.
No unit for channel input mode (
Example:
//Position marker 2 to frequency 10.7 MHz
CALC:MARK2:X 10.7MHz
Manual operation:
See
See
See
See
CALCulate<n>:MARKer<m>:X:SLIMits<slim>:LEFT
<SearchLimit>
This command defines the left limit of the marker search range.
Prerequisites for this command
●
Turn on marker search limits (
If you perform a measurement in zero span, this command limits the evaluation range
on the trace.