Remote Control
R&S
®
FPC
282
User Manual 1178.4130.02 ─ 07
19.12.9
Marker Functions
CALCulate<n>:MARKer<m>:COUNt:FREQuency?
............................................................282
CALCulate<n>:MARKer<m>:COUNt[:STATe]
.................................................................... 282
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:HOLD
............................................. 283
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:SELect
............................................283
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation[:STATe]
........................................... 284
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown
............................................................ 284
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:FREQuency?
........................................285
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:RESult?
............................................... 285
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:STATe
.................................................. 286
CALCulate<n>:MARKer<m>:FUNCtion:NOISe:RESult?
.....................................................286
CALCulate<n>:MARKer<m>:FUNCtion:NOISe[:STATe]
......................................................287
................................................................................................287
CALCulate<n>:MARKer<m>:COUNt:FREQuency?
This command measures the frequency at the marker position and returns the result.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end to make sure that the R&S
FPC actually reaches the frequency you want to
measure. This is only possible in single sweep mode.
Prerequisites for this command
●
Turn on the frequency counter (
CALCulate<n>:MARKer<m>:COUNt[:STATe]
).
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Return values:
<Frequency>
<numeric value>
Default unit: Hz
Example:
See
CALCulate<n>:MARKer<m>:COUNt[:STATe]
Usage:
Query only
Manual operation:
See
CALCulate<n>:MARKer<m>:COUNt[:STATe]
<State>
This command turns the frequency counter at the marker position on and off.
You can read out the result with
CALCulate<n>:MARKer<m>:COUNt:FREQuency?
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end to make sure that the R&S
FPC actually reaches the frequency you want to
measure. This is only possible in single sweep mode.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Remote Commands of the Spectrum Application