Preparing for Use
R&S
®
FPC
14
User Manual 1178.4130.02 ─ 07
Turning on the R&S
FPC
► Turn on the main AC power switch on the rear panel of the R&S
FPC (position "I").
The instrument is now supplied with AC power.
● "Power" key is highlighted orange: R&S
FPC is in standby mode (main AC
power switch is in position "I").
● "Power" key is highlighted green: R&S
FPC is running and ready for operation.
Turning off the R&S
FPC
► Turn off the main AC power switch on the rear panel of the R&S
FPC (position
"O").
The instrument is no longer supplied with AC power.
Changing the AC supply fuse
Only fuses of the type 2A T IEC60127-2/V should be used.
1. Disconnect the power cable.
2. Open the flap covering the voltage selector using a small screwdriver (or similar).
3. Remove the cylinder labeled with the nominal voltages. Remove the fuse and
install the new one. Reinsert the cylinder so that the value visible through the hole
in the cover flap is the same nominal voltage as before.
4. Close the flap.
Turning the R&S
FPC on and off